E2075/E2075M-10 Standard Practice for Verifying the Consistency of AE-Sensor Response Using an Acrylic Rod
Published By | Publication Date | Number of Pages |
ASTM | 2010 | 4 |
1.1 This practice is used for routinely checking the sensitivity of acoustic emission (AE) sensors. It is intended to provide a reliable, precisely specified way of comparing a set of sensors, or telling whether an individual sensor’s sensitivity has degraded during its service life, or both.
1.2 This practice is not a calibration nor does it give frequency response information.
1.3 Units The values stated in SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard.
1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.