D6443-04(2010) Test Method for Determination of Calcium, Chlorine, Copper, Magnesium, Phosphorus, Sulfur, and Zinc in Unused Lubricating Oils and Additives by Wavelength Dispersive X-ray Fluorescence Spectrometry (Mathematical Correction Procedure)
Published By | Publication Date | Number of Pages |
ASTM | 2010 | 7 |
ASTM D6443-04-Reapproved2010
Historical Standard: Test Method for Determination of Calcium, Chlorine, Copper, Magnesium, Phosphorus, Sulfur, and Zinc in Unused Lubricating Oils and Additives by Wavelength Dispersive X-ray Fluorescence Spectrometry (Mathematical Correction Procedure)
ASTM D6443
Scope
1.1 This test method covers the determination of calcium, chlorine, copper, magnesium, phosphorus, sulfur, and zinc in unused lubricating oils, additives, and additive packages by wavelength dispersive X-ray fluorescence spectrometry. Matrix effects are handled with mathematical corrections.
1.2 For each element, the upper limit of the concentration range covered by this test method is defined by the highest concentration listed in Table 1. Samples containing higher concentrations can be analyzed following dilution.
1.3 For each element, the lower limit of the concentration range covered by this test method can be estimated by the limit of detection (LOD) (see also 40 CFR 136 Appendix B) or limit of quantification (LOQ), both of which can be estimated from Sr, the repeatability standard deviation. LOD and LOQ values, determined from results obtained in the interlaboratory study on precision, are listed in Table 2.
1.3.1 LOD and LOQ are not intrinsic constants of this test method. LOD and LOQ depend upon the precision attainable by a laboratory when using this test method.
1.4 This test method uses regression software to determine calibration parameters, which can include influence coefficients (that is, interelement effect coefficients) (Guide E1361), herein referenced as alphas. Alphas can also be determined from theory using relevant software.
1.5 Setup of this test method is intended for persons trained in the practice of X-ray spectrometry. Following setup, this test method can be used routinely.
1.6 The values stated in either SI units or angstrom units are to be regarded separately as standard.
1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.
Keywords
additive elements; additives; calcium; chlorine; copper; lubricating oils; magnesium; phosphorus; sulfur; wavelength dispersive X-ray spectrometry; zinc; Additives–lubricating oils; Additives–petroleum products; Calcium content–petroleum products; Calibration–petroleum analysis instrumentation; Chlorine content–petroleum products; Copper content–petroleum products; Fuel sulfur; Magnesium content–petroleum products; Phosphorus content–petroleum products; Wavelength dispersive X-ray fluorescence (WDXRF); XRF (X-ray fluorescence)–petroleum products; Zinc content–petroleum products
ICS Code
ICS Number Code 75.100 (Lubricants, industrial oils and related products)
DOI: 10.1520/D6443-04R10
PDF Catalog
PDF Pages | PDF Title |
---|---|
1 | Scope Referenced Documents TABLE 1 |
2 | Summary of Test Method Significance and Use Interferences Apparatus TABLE 2 |
3 | Reagents and Materials Sampling and Sample Handling Preparation of Calibration Standards Calibration TABLE 3 |
4 | Analysis Quality Control (Required) Reporting TABLE 4 |
5 | Precision and Bias TABLE 5 TABLE 6 TABLE 7 |
6 | Keywords X1. AIDS TO THE ANALYST X1.1 X1.2 X1.3 X1.4 X1.5 X1.6 X1.7 X1.8 X1.9 X1.10 X1.11 X1.12 X1.13 X1.14 X1.15 TABLE 8 |
7 | X2. QUALITY CONTROL (QC) MONITORING X2.1 X2.2 X2.3 X2.4 X2.5 X2.6 |